Salt spray test chamber common faults and treatment methods

Because the salt spray tester has long been tested for salt spray aging, the environment is prone to cause the salt spray test machine to have large and small faults. In order to solve the fault quickly, the salt spray aging test will continue smoothly. Some common faults are described below. Approach.

Saturated barrel temperature cannot rise to the set temperature

cause of issue:

1. The temperature of the saturated barrel temperature controller is too low;

2. The saturated barrel safety protection switch setting is too low;

3. Heating system failure;

4. Electromagnetic relay failure;

5. Controller failure.

Approach:

1. Set the temperature controller to the desired temperature;

2. Set the safety switch to the desired temperature;

Insufficient fog

cause of issue:

1. The spray regulator is placed too low;

2. The glass filter in the preheating tank is blocked;

3. The pressure setting is too low.

Approach:

1. Raise the spray regulator;

2. Clean the glass filter;

3. Adjust the pressure regulating valve to a pressure of 1Kg per square centimeter. The air compressor is marked with a pressure regulating valve and adjusted to a pressure of 2Kg per square centimeter.

Inadequate water level

Cause: The water level is too low.

Treatment: Check if the test chamber or saturated tank is adequately water level.

Air compressor

The cause of the malfunction: The air compressor itself has a self-protection function.

Treatment: Use as usual without affecting the test process.

Unable to spray

cause of issue:

1. The air compressor is not running;

2. The main switch of the air compressor outlet is not open;

3. Solenoid valve failure;

4. The pressure gauge is faulty or the pressure is too low;

5. Electromagnetic contactor failure;

Nozzle blockage

Approach:

1. Turn on the air compressor button;

2. Turn on the air compressor main switch;

3. Remove the nozzle and clean it (please disassemble it carefully).

Unable to operate

Cause of failure: When the water level in the heating tank is too low, the power supply will be cut off.

Treatment method: Add the water level of the heating tank to the normal condition.

thermostat

The cause of the malfunction: the temperature controller is faulty and the temperature sensing rod is faulty.

Treatment method: notify the maintenance company to repair the door.